CHAPTER 10
QUALITY CONTROL FOR TESTS OF MODERATE COMPLEXITY TESTING (INCLUDING THE
SUBCATEGORY), HIGH COMPLEXITY TESTING OR ANY COMBINATION OF THESE TESTS
10-1. Condition: General quality control; moderate complexity testing
(including the subcategory), high complexity testing, or any combination
of these tests (§493.1201)
a. Chapter 10 is divided into two sections, general quality control and
quality control for specialties and subspecialties. The quality control
requirements are specified in paragraph 10-1 through paragraph 10-43 unless:
(1) An alternate procedure specified in the manufacturer's protocol has been cleared by the Food and Drug Administration (FDA) as meeting certain CLIA requirements for general quality control and specialty/subspecialty quality control, and the manufacturer's instructions contain the following statement,
"Unless this device is modified by a laboratory, the laboratory's compliance with these quality control instructions will satisfy the applicable requirements of 42 CFR paragraph 1203b."; or
(2) OASD(HA) approves an equivalent procedure that is specified in Appendix C of the State Operations Manual (HCFA Pub. 7). The DoD CLIAC will coordinate/notify HCFA and CDC for all modifications made to CLIA regulations,
b. The laboratory must establish and follow written quality control procedures
for monitoring and evaluating the quality of the analytical testing process
of each method to assure the accuracy and reliability of patient test results
and reports. The laboratory must meet the applicable standards in paragraphs
10-2 through 10-12, unless an alternative procedure specified in the manufacturer's
protocol has been cleared by the Food and Drug Administration (FDA) as meeting
certain CLIA requirements for quality control or OASD(HA) approves an equivalent
procedure specified in appendix C of the State Operations Manual (HCFA Pub.
7). HCFA Pub. 7 is available from the Technical Information Service, U.S.
Department of Commerce, 5825 Port Royal Road, Springfield, VA 22161,telephone
number (703) 487-4630.
10-2. Standard: Moderate complexity testing (including the subcategory),
high complexity testing, or any combination of these tests: Effective 1
September 1993 to 1 September 1996 (§493.1202)
a. For each test of high complexity performed, the laboratory must meet
all applicable standards of this chapter.
b. For each test of moderate complexity performed using a standardized method
or method developed in-house, a device not subject to clearance by the FDA
(including any commercially distributed instrument, kit or test system subject
to the Food and Drug Cosmetic Act prior to the Medical Device Amendments,
Public Law 94-295, enacted on 28 May 1976, and those identified in 21 CFR
parts 862, 864, and 866 as exempt from FDA premarket review) or using an
instrument, kit or test system cleared by the FDA through the premarket
notification (510(k)) or premarket approval (PMA) process for in-vitro diagnostic
use but modified by the laboratory, the laboratory must meet all applicable
standards of this chapter.
c. For all other tests of moderate complexity performed using an instrument,
kit or test system cleared by the FDA through the premarket notification
(510(k)) or premarket approval (PMA) process for in-vitro diagnostic use,
the laboratory must:
(1) Follow the manufacturer's instructions for instrument or test system operation and test performance.
(2) Have a procedure manual describing the processes for testing and reporting patient test results.
(3) Perform and document calibration procedures, or check calibration at least once every six months.
(4) Perform and document control procedures using at least two levels of control materials each day of testing.
(5) Perform and document applicable specialty and subspecialty control procedures as specified under paragraph 10-13.
(6) Perform and document that remedial action has been taken when problems or errors are identified as specified in paragraph 10-11.
(7) Maintain records of all quality control activities for two years. Quality control records for immunohematology and blood and blood products must be maintained as specified in paragraph 10-12.
10-3. Standard: Moderate Complexity (including the subcategory), high
complexity testing, or any combination of these tests (effective beginning
1 July 1998) (§493.1203).
For each moderate or high complexity test performed, the laboratory will
be in compliance with this section if it:
a. Meets all applicable quality control requirements specified in this chapter
when using a standardized method, a method developed in-house, a device
not subject to clearance by the FDA (including any commercially distributed
instrument, kit or test system subject to the Food, Drug and Cosmetic Act
prior to the Medical Device Amendments, Public Law 94-295, enacted on 28
May 1976, and those identified in 21 CFR parts 862, 864 and 866 as exempt
from FDA premarket review), a manufacturer's product modified by the laboratory,
or a device (instrument, kit or test system) not cleared by the FDA as meeting
certain CLIA quality control requirements; or
b. Follows manufacturer's instructions when using a device (instruments,
kits, or test systems) cleared by the FDA as meeting the CLIA requirements
for general quality control located at paragraphs 10-8, 10-9, and 10-13,
and applicable parts of paragraphs 10-5, 10-6, and 10-10. In addition, the
laboratory must comply with the requirements of paragraphs 10-4, 10-7, 10-11,
and 10-12, and those parts of paragraphs 10-5, 10-6 and 10-10 that are unique
to the laboratory facility and cannot be met by following manufacturer's
instructions.
10-4. Standard: Facilities (§493.1204)
The laboratory must provide the space and environmental conditions necessary
for conducting the services offered.
a. The laboratory must be constructed, arranged, and maintained to ensure
the space, ventilation, and utilities necessary for conducting all phases
of testing, including the preanalytic (pre-testing), analytic (testing),
and postanalytic (post-testing), as appropriate.
b. Safety precautions must be established, posted, and observed to ensure
protection from physical, chemical, biochemical and electrical hazards and
biohazardous materials.
10-5. Standard: Test methods, equipment, instrumentation, reagents, materials,
and supplies (§493.1205)
The laboratory must utilize test methods, equipment, instrumentation, reagents,
materials, and supplies that provide accurate and reliable test results
and test reports.
a. Test methodologies and equipment must be selected and testing performed
in a manner that provides test results within the laboratory's stated performance
specifications for each test method as determined under paragraph 10-7.
b. The laboratory must have appropriate and sufficient equipment, instruments,
reagents, materials, and supplies for the type and volume of testing performed
and for the maintenance of quality during the preanalytic, analytic, and
postanalytic phases of testing.
c. The laboratory must define criteria for those conditions that are essential
for proper storage of reagents and specimens, and accurate and reliable
test system operation and test result reporting.
(1) These conditions include, if applicable; water quality, temperature, humidity, and protection of equipment and instrumentation from fluctuations and interruptions in electrical current that adversely affect patient test results and test reports.
(2) Remedial actions taken to correct conditions that fail to meet the criteria specified in para-graph c(1) above must be documented.
d. Reagents, solutions, culture media, control materials, calibration materials
and other supplies, as appropriate, must be labeled to indicate:
e. Reagents, solutions, culture media, control materials, calibration materials
and other supplies must be prepared, stored, and handled in a manner to
ensure that:
(1) Reagents, solutions, culture media, controls, calibration materials and other supplies are not used when they have exceeded their expiration date, have deteriorated or are of substandard quality. The laboratory must comply with the FDA product dating requirements of 21 CFR Part 610.53 for blood products and other biologicals, and labeling requirements, as cited in 21 CFR Part 809.10, for all other in vitro diagnostics. Any exception to the product dating requirements in 21 CFR Part 610.53 will be granted by the FDA in the form of an amendment of the product license, in accordance with 21 CFR Part 610.53d. All exceptions must be documented by the laboratory.
(2) Components of reagent kits of different lot numbers are not interchanged unless otherwise specified by the manufacturer.
10-6. Standard: Procedure manual (§493.1211)
a. A written procedure manual for the performance of all analytical methods
used by the laboratory must be readily available and followed by laboratory
personnel. Textbooks may be used as supplements to these written descriptions
but may not be used in lieu of the laboratory's written procedures for testing
or examining specimens.
b. The procedure manual must include, when applicable to the test procedure:
(1) Requirements for specimen collection and processing, and criteria for specimen rejection.
(2) Procedures for microscopic examinations, including the detection of inadequately prepared slides.
(3) Step-by-step performance of the procedure, including test calculations and interpretation of results.
(4) Preparation of slides, solutions, calibrators, controls, reagents, stains and other materials used in testing.
(5) Calibration and calibration verification procedures.
(6) The reportable range for patient test results as established or verified in paragraph 10-7 below.
(7) Control procedures.
(8) Remedial action to be taken when calibration or control results fail to meet the laboratory's criteria for acceptability.
(9) Limitations in methodologies, including interfering substances.
(10) Reference range (normal values).
(11) Imminent life-threatening laboratory results or "panic values".
(12) Pertinent literature references.
(13) Appropriate criteria for specimen storage and preservation to ensure specimen integrity until testing is completed.
(14) The laboratory's system for reporting patient results including, when appropriate, the protocol for reporting panic values.
(15) Description of the course of action to be taken in the event that a test system becomes inoperable.
(16) Criteria for the referral of specimens including procedures for specimen submission and handling as described in paragraph 9-2.
c. Manufacturers' package inserts or operator manuals may be used, when
applicable, to meet the requirements of paragraphs b(1) through b(13) above.
Any of the items under paragraphs b(1) through b(13) above not provided
by the manufacturer must be provided by the laboratory.
d. Procedures must be approved, signed, and dated by the director.
e. Procedures must be re-approved, signed and dated if the directorship
of the laboratory changes.
f. Each change in a procedure must be approved, signed, and dated by the
current laboratory director.
g. The laboratory must maintain a copy of each procedure with the dates
of initial use and discontinuance. These records must be retained for two
years after a procedure has been discontinued.
10-7. Standard: Establishment and verification of method performance
specifications (§493.1213)
Prior to reporting patient test results, the laboratory must verify or establish,
for each method, the performance specifications for the following performance
characteristics: accuracy; precision; analytical sensitivity and specificity,
if applicable; the reportable range of patient test results; the reference
range(s) (normal values); and any other applicable performance characteristic.
a. The provisions of this chapter are not retroactive. Laboratories are
not required to verify or establish performance specifications for any test
method of moderate or high complexity in use prior to 1 September 1992.
b. Each laboratory that introduces a new procedure for patient testing using
a device (instrument, kit or test system) cleared by the FDA meeting certain
CLIA requirements for quality, control must demonstrate that, prior to reporting
patient test results, it can obtain the performance specifications for accuracy,
precision, and reportable range of patient test results, comparable to those
established by the manufacturer. The laboratory must also verify that the
manufacturer's reference range is appropriate for the laboratory's patient
population.
c. Each laboratory that introduces a new procedure or device as specified
in either paragraph 10-2a or b, or paragraph 10-3a above, must, prior to
reporting test results:
(1) Verify or establish for each method the performance specifications for the following performance characteristics, as applicable:
(a) Accuracy.
(b) Precision.
(c) Analytical sensitivity.
(d) Analytical specificity to include interfering substances.
(e) Reportable range of patient test results.
(f) Reference range(s).
(g) Any other performance characteristic required for test performance.
(2) Based upon the performance specifications verified or established in accordance with paragraph c(1) above, establish calibration and control procedures for patient testing as required under paragraphs 10-9 and 10-10 below.
d. The laboratory must have documentation of the verification or establishment
of all applicable test performance specifications.
10-8. Standard: Equipment maintenance and function checks (§493.1215)
The laboratory must perform equipment maintenance and function checks that
include electronic, mechanical and operational checks necessary for the
proper test performance and test result reporting of equipment, instruments
and test systems, to assure accurate and reliable test results and reports.
a. Maintenance of equipment, instruments, and test systems.
(1) For manufacturers' equipment, instruments or test systems cleared by the FDA as meeting the CLIA requirements for quality control, the laboratory must:
(a) Perform maintenance as defined by the manufacturer and with at least the frequency specified by the manufacturer.
(b) Document all maintenance performed.
(2) For methods or devices as specified in either paragraph 10-2a or b or paragraph 10-3a above, the laboratory must:
(a) Establish a maintenance protocol that ensures equipment, instrument, and test system performance necessary for accurate and reliable test results and test result reporting.
(b) Perform maintenance with at least the frequency specified in paragraph a(2)(a) above.
(c) Document all maintenance performed.
b. Function checks of equipment, instruments, and test systems.
(1) For manufacturers' equipment, instruments, or test systems cleared by the FDA as meeting the CLIA requirements for quality control, the laboratory must:
(a) Perform function checks as defined by the manufacturer and with at least the frequency specified by the manufacturer.
(b) Document all function checks performed.
(2) For methods or devices as specified in either paragraph 10-2a or b or paragraph 10-3a above, the laboratory must:
(a) Define a function check protocol that ensures equipment, instrument, and test system performance necessary for accurate and reliable test results and test result reporting.
(b) Perform function checks including background or baseline checks specified in paragraph b(2)(a) above. Function checks must be within the laboratory's established limits before patient testing is conducted.
(c) Document all function checks performed.
10-9. Standard: Calibration and calibration verification procedures (§493.1217)
Calibration and calibration verification procedures are required to substantiate
the continued accuracy of the test method throughout the laboratory's reportable
range for patient test results. Calibration is the process of testing and
adjusting an instrument, kit, or test system to provide a known relationship
between the meas-urement response and the value of the substance that is
being measured by the test procedure. Calibration verification is the assaying
of calibration materials in the same manner as patient samples to confirm
that the calibration of the instrument, kit, or test system has remained
stable throughout the laboratory's report-able range for patient test results.
The reportable range is the range of test result values over which the relationship
between the instrument, kit or test system measurement response is shown
to be valid. The reportable range of patient test results is the range of
test result values over which the laboratory can establish or verify the
accuracy of the instrument, kit or test system measurement response. Calibration
and calibration verification must be performed and documented as required
in this section unless otherwise specified in paragraphs 10-13 through 10-43.
a. For laboratory test procedures that are performed using instruments,
kits, or test systems that have been cleared by the FDA as meeting certain
CLIA requirements for quality control, the laboratory must, at a minimum,follow
the manufacturer's instructions for calibration and calibration verification
procedures using calibration materials specified by the manufacturer.
b. For each method or device, as specified in either paragraph 10-2a, 10-2b
or paragraph 10-3, the laboratory must:
(1) Perform calibration procedures:
(a) At a minimum, in accordance with manufacturer's instructions, if provided, using calibration materials provided or specified, as appropriate, and with at least the frequency recommended by the manufacturer.
(b) In accordance with criteria established by the laboratory, as required under paragraph 10-7c(1) including the number, type and concentration of calibration materials, acceptable limits for calibration, and the frequency of calibration if manufacturer's instructions are not provided and using calibration materials appropriate for the methodology and, if possible, traceable to a reference method or reference material of known value.
(c) Whenever calibration verification fails to meet the laboratory's acceptable limits for calibration verification; and
(2) Perform calibration verification procedures:
(a) In accordance with the manufacturer's calibration verification instructions when they meet or exceed the requirements specified in paragraph b(2)(b) below.
(b) In accordance with criteria established by the laboratory including the number, type, and concentration of calibration materials, acceptable limits for calibration verification and frequency of calibration verification using calibration materials appropriate for the methodology and, if possible, traceable to a reference method or reference material of known value and verifying the laboratory's established reportable range of patient test results, which must include at least a minimal (or zero) value, a mid-point value, and a maximum value at the upper limit of that range.
(c) At least once every six months and whenever any of the following occur:
- A complete change of reagents for a procedure is introduced, unless the laboratory can demonstrate that changing reagent lot numbers does not affect the range used to report patient test results, and control values are not adversely affected by reagent lot number changes. NOTE: If reagents are obtained from a manufacturer and all of the reagents for a test are packaged together, the laboratory is not required to perform calibration verification for each package of reagents, provided the packages of reagents are received in the same shipment and contain the same lot number.
- There is major preventive maintenance or replacement of critical parts that may influence test performance.
- Controls reflect an unusual trend or shift or are outside of the laboratory's acceptable limits and other means of assessing and correcting unacceptable control values have failed to identify and correct the problem.
- The laboratory's established schedule for verifying the reportable range for patient test results requires more frequent calibration verification than specified in paragraphs b(2)(c).
(3) Document all calibration and calibration verification procedures performed.
10-10. Standard: Control procedures (§493.1218)
Control procedures are performed on a routine basis to monitor the stability
of the method or test system; control and calibration materials provide
a means to indirectly assess the accuracy and precision of patient test
results. Control procedures must be performed as defined in this section
unless otherwise specified in paragraphs 10-13 through 10-43.
a. For each method or device cleared by the FDA as meeting CLIA requirements
for general quality control, the laboratory must, at a minimum, follow the
manufacturer's instructions for control procedures. In addition, the laboratory
must meet the requirements under paragraphs c through e below and, as applicable,
paragraph f below.
b. For each device as specified in paragraph 10-2a or b or paragraph 10-3a
above, the laboratory must evaluate instrument and reagent stability and
operator variance in determining the number, type, and frequency of testing
calibration or control materials and establish criteria for acceptability
used to monitor test performance during a run of patient specimen(s). A
run is an interval within which the accuracy and precision of a testing
system is expected to be stable, but cannot be greater than 24 hours or
less than the frequency recommended by the manufacturer. For each procedure,
the laboratory must monitor test performance using calibration materials
or control materials or a combination thereof.
(1) For qualitative tests, the laboratory must include a positive and negative control with each run of patient specimens.
(2) For quantitative tests, the laboratory must include at least two samples of different concen-trations of either calibration materials, control materials, or a combination thereof with the frequency determined in paragraph 10-10b, but not less frequently than once each run of patient specimens.
(3) For electrophoretic determinations:
(a) At least one control sample must be used in each electrophoretic cell.
(b) The control sample must contain fractions representative of those routinely reported in patient specimens.
(4) Each day of use, the laboratory must evaluate the detection phase of direct antigen systems using an appropriate positive and negative control material (organism or antigen extract). When direct antigen systems include an extraction phase, the system must be checked each day of use using a positive organism.
(5) If calibration materials and control materials are not available, the laboratory must have an alternative mechanism to assure the validity of patient test results.
c. Control samples must be tested in the same manner as patient specimens.
d. When calibration or control materials are used, statistical parameters
(e.g., mean and standard deviation) for each lot number of calibration material
and each lot of control material must be determined through repetitive testing.
(1) The stated values of an assayed control material may be used as the target values provided the stated values correspond to the methodology and instrumentation employed by the laboratory and are verified by the laboratory.
(2) Statistical parameters for unassayed materials must be established over time by the laboratory through concurrent testing with calibration materials or control materials having previously determined statistical parameters.
e. Control results must meet the laboratory's criteria for acceptability
prior to reporting patient test results.
f. Reagent and supply checks.
(1) The laboratory must check each batch or shipment of reagents, discs, stains, antisera and identification systems (using two or more substrates) when prepared or opened for positive and negative reactivity, as well as graded reactivity if applicable.
(2) Each day of use (unless otherwise specified in this chapter), the laboratory must test staining materials for intended reactivity to ensure predictable staining characteristics.
(3) The laboratory must check fluorescent stains for positive and negative reactivity each time of use (unless otherwise specified in this chapter).
(4) The laboratory must check each batch or shipment of media for sterility, if it is intended to be sterile, and sterility is required for testing. Media must also be checked for its ability to support growth, and as appropriate, selectivity/inhibition and/or biochemical response. The laboratory may use manufacturer's control checks of media provided the manufacturer's product insert specifies that the manufacturer's quality control checks meet the National Committee for Clinical Laboratory Standards (NCCLS) for media quality control. The laboratory must document that the physical characteristics of the media are not compromised and report any deterioration in the media to the manufacturer. The laboratory must follow the manufacturer's specifications for using the media and be responsible for the test results. NOTE: A batch of media (solid, semi-solid, or liquid) consists of all tubes, plates, or containers of the same medium prepared at the same time and in the same laboratory; or, if received from an outside source or commercial supplier, consists of all of the plates, tubes or containers of the same medium that have the same lot numbers and are received in a single shipment.
10-11. Standard: Remedial actions (§493.1219)
Remedial action policies and procedures must be established by the laboratory
and applied as necessary to maintain the laboratory's operation for testing
patient specimens in a manner that assures accurate and reliable patient
test results and reports. The laboratory must document all remedial actions
taken when:
a. Test systems do not meet the laboratory's established performance specifications,
as determined in paragraph 10-7 above, which includes, but are not limited
to:
(1) Equipment or methodologies that perform outside of established operating parameters or performance specifications,
(2) Patient test values that are outside of the laboratory's reportable range of patient test results.
(3) The determination that the laboratory's reference range for a test procedure is inappropriate for the laboratory's patient population.
b. Results of control and calibration materials fail to meet the laboratory's
established criteria for acceptability. All patient test results obtained
in the unacceptable test run or since the last acceptable test run must
be evaluated to determine if patient test results have been adversely affected
and the laboratory must take the remedial action necessary to ensure the
reporting of accurate and reliable patient test results.
c. The laboratory cannot report patient test results within its established
time frames. The laboratory must determine, based on the urgency of the
patient test(s) requested, the need to notify the appropriate individual
of the delayed testing.
d. Errors in the reported patient test results are detected. The laboratory
must:
(1) Promptly notify the authorized person ordering or individual utilizing the test results of reporting errors.
(2) Issue corrected reports promptly to the authorized person ordering the test or the individual utilizing the test results.
(3) Maintain exact duplicates of the original report as well as the corrected report for two years.
10-12. Standard: Quality control records (§493.1221)
The laboratory must document and maintain records of all quality control
activities specified in paragraphs 10-2 through 10-43 and retain records
for at least two years. Immunohematology quality control records must be
maintained for a period of no less than five years. In addition, quality
control records for blood and blood products must be maintained for a period
not less than five years after processing records have been completed, or
six months after the latest expiration date, whichever is the later date,
in accordance with 21 CFR Part 606.160d.
10-13. Condition: Quality control; Specialties and subspecialties for
tests of moderate or high complexity, or both (§493.1223)
The laboratory must establish and follow written quality control procedures
for monitoring and evaluating the quality of the analytical testing process
of each method to assure the accuracy and reliability of patient test results
and reports. Except as specified in paragraph 10-2c above, the laboratory
must meet the applicable general requirements specified in paragraphs 10-1
through 10-12. In addition, the laboratory must meet the applicable requirements
of paragraphs 10-14 through 10-43 below unless an alternative procedure
specified in the manufacturer's protocol has been cleared by the Food and
Drug Administration (FDA) as meeting certain CLIA requirements for quality
control or OASD(HA) approves an equivalent procedure. Failure to meet any
of the applicable conditions in paragraphs 10-14 through 10-43 below will
result in intermediate sanctions, and/or revocation of CLIP certification
for the entire specialty or subspecialty to which the condition applies,
in accordance with Chapter 14.
10-14. Condition: Microbiology (§493.1225)
The laboratory must meet the applicable quality control requirements in
paragraphs 10-1 through 10-12 and paragraphs 10-15 through 10-19 for the
subspecialties for which it is certified under the specialty of microbiology.
10-15. Condition: Bacteriology (§493.1227)
To meet the quality control requirements for bacteriology, the laboratory
must comply with the applicable requirements in paragraphs 10-1 through
10-12 above and with paragraphs a through c below. All quality control activities
must be documented.
a. The laboratory must check positive and negative reactivity with control
organisms:
(1) Each day of use for catalase, coagulase, betalactamase, and oxidase reagents and DNA probes.
(2) Each week of use for Gram and acid-fast stains, bacitracin, optochin, ONPG, X, and V discs or strips.
(3) Each month of use for antisera.
b. Each week of use, the laboratory must check XV discs or strips with a
positive control organism.
c. For antimicrobial susceptibility tests, the laboratory must check each
new batch of media and each lot of antimicrobial discs before, or concurrent
with, initial use, using approved reference organisms.
(1) The laboratory's zone sizes or minimum inhibitory concentration for reference organisms must be within established limits before reporting patient results.
(2) Each day tests are performed, the laboratory must use the appropriate control organism(s) to check the procedure.
10-16. Condition: Mycobacteriology (§493.1229)
To meet the quality control requirements for mycobacteriology, the laboratory
must comply with the applicable requirements in paragraphs 10-1 through
10-12 above and with paragraphs a through d below. All quality control activities
must be documented.
a. Each day of use, the laboratory must check the iron uptake test with
at least one acid-fast organism that produces a positive reaction and with
an organism that produces a negative reaction and check all other reagents
or test procedures used for mycobacteria identification with at least one
acid-fast organism that produces a positive reaction.
b. The laboratory must check fluorochrome acid-fast stains for positive
and negative reactivity each week of use.
c. The laboratory must check acid-fast stains each week of use with an acid-fast
organism that produces a positive reaction.
d. For susceptibility tests performed on Mycobacterium tuberculosis
isolates, the laboratory must check the procedure each week of use with
a strain of Mycobacterium tuberculosis susceptible to all antimycobacterial
agents tested.
10-17. Condition: Mycology (§493.1231)
To meet the quality control requirements for mycology, the laboratory must
comply with the applicable requirements in paragraphs 10-1 through 10-12
above and with paragraphs a through d below. All quality control activities
must be documented.
a. Each day of use, the laboratory using the auxanographic medium for nitrate
assimilation must check the nitrate reagent with a peptone control.
b. Each week of use, the laboratory must check all reagents used with biochemical
tests and other test procedures for mycological identification with an organism
that produces a positive reaction.
c. Each week of use, the laboratory must check acid-fast stains for positive
and negative reactivity.
d. For susceptibility tests, the laboratory must test each drug each day
of use with at least one control strain that is susceptible to the drug.
The laboratory must establish control limits. Criteria for acceptable control
results must be met prior to reporting patient results.
10-18. Condition: Parasitology (§493.1233)
To meet the quality control requirements for parasitology, the laboratory
must comply with the applicable requirements of paragraphs 10-1 through
10-12 above and with paragraphs a through c below. All quality control activities
must be documented.
a. The laboratory must have available a reference collection of slides or
photographs, and, if available, gross specimens for identification of parasites
and use these references in the laboratory for appropriate comparison with
diagnostic specimens.
b. The laboratory must calibrate and use the calibrated ocular micrometer
for determining the size of ova and parasites, if size is a critical parameter.
c. Each month of use, the laboratory must check permanent stains using a
fecal sample control that will demonstrate staining characteristics.
10-19. Condition: Virology (§493.1235)
To meet the quality control requirements for virology, the laboratory must
comply with the applicable requirements in paragraphs 10-1 through 10-12
above and with paragraphs a through c below. All quality control activities
must be documented.
a. The laboratory must have available host systems for the isolation of
viruses and test methods for the identification of viruses that cover the
entire range of viruses that are etiologically related to clinical diseases
for which services are offered.
b. The laboratory must maintain records that reflect the systems used and
the reactions observed.
c. In tests for the identification of viruses, the laboratory must simultaneously
culture uninoculated cells or cell substrate controls as a negative control
to detect erroneous identification results.
10-20. Condition: Diagnostic immunology (§493.1237)
The laboratory must meet the applicable quality control requirements in
paragraphs 10-1 through 10-12 above and paragraphs 10-21 and 10-22 below
for the subspecialties for which it is certified under the specialty of
diagnostic immunology.
10-21. Condition: Syphilis serology (§493.1239)
To meet the quality control requirements for syphilis serology, the laboratory
must comply with the applicable requirements in paragraphs 10-1 through
10-12 above and with paragraphs a through e below. All quality control activities
must be documented.
a. For laboratories performing syphilis testing, the equipment, glassware,
reagents, controls, and techniques for tests for syphilis must conform to
manufacturers' specifications.
b. The laboratory must run serologic tests on patient specimens concurrently
with a positive serum control of known titer or controls of graded reactivity
plus a negative control.
c. The laboratory must employ positive and negative controls that evaluate
all phases of the test system to ensure reactivity and uniform dosages.
d. The laboratory may not report test results unless the predetermined reactivity
pattern of the controls is observed.
e. All facilities manufacturing blood and blood products for transfusion
or serving as referral laboratories for these facilities must meet the syphilis
serology testing requirements of 21 Part CFR 640.5(a).
10-22. Condition: General immunology (§493.1241)
To meet the quality control requirements for general immunology, the laboratory
must comply with the applicable requirements in paragraphs 10-1 through
10-12 above and with paragraphs a through d below. All quality control activities
must be documented.
a. The laboratory must run serologic tests on patient specimens concurrently
with a positive serum control of known titer or controls of graded reactivity,
if applicable, plus a negative control.
b. The laboratory must employ controls that evaluate all phases of the test
system (antigens, complement, erythrocyte indicator systems, etc.) to ensure
reactivity and uniform dosages when positive and negative controls alone
are not sufficient.
c. The laboratory may not report test results unless the predetermined reactivity
pattern of the controls is observed.
d. All facilities manufacturing blood and blood products for transfusion
or serving as referral laboratories for these facilities must meet:
(1) The HIV testing requirements of 21 CFR Part 610.45; and
(2) Hepatitis testing requirements of 21 CFR Part 610.40.
10-23. Condition: Chemistry (§493.1243)
The laboratory must meet the applicable quality control requirements in
paragraphs 10-1 through 10-12 above and paragraphs 10-24 through 10-26 below
for the subspecialties for which it is certified under the specialty of
chemistry. All quality control activities must be documented.
10-24. Condition: Routine chemistry (§493.1245)
To meet the quality control requirements for routine chemistry, the laboratory
must comply with the applicable requirements in paragraphs 10-1 through
10-12 above. All quality control activities must be documented. In addition,
for blood gas analyses, the laboratory must:
a. Calibrate or verify calibration according to the manufacturer's specifications
and with at least the frequency recommended by the manufacturer.
b. Test one sample of control material each eight hours of testing.
c. Use a combination of calibrators and control materials that include both
low and high values on each day of testing.
d. Include one sample of calibration material or control material each time
patients are tested unless automated instrumentation internally verifies
calibration at least every thirty minutes.
10-25. Condition: Endocrinology (§493.1247)
To meet the quality control requirements for endocrinology, the laboratory
must comply with the applicable requirements contained in paragraphs 10-1
through 10-12 above. All quality control activities must be documented.
10-26. Condition: Toxicology (§493.1249)
To meet the quality control requirements for toxicology, the laboratory
must comply with the applicable requirements in paragraphs 10-1 through
10-12 above. All quality control activities must be documented. In addition,
for drug abuse screening using thin layer chromatography:
a. Each plate must be spotted with at least one sample of calibration material
containing all drug groups identified by thin layer chromatography which
the laboratory reports.
b. At least one control sample must be included in each chamber, and the
control sample must be processed through each step of patient testing, including
extraction procedures.
10-27. Condition: Urinalysis (§493.1251)
Except in those tests categorized as minimal, to meet the quality control
requirements for urinalysis, the laboratory must comply with the applicable
requirements in paragraphs 10-1 through 10-12 above. All quality control
activities must be documented.
10-28. Condition: Hematology (§493.1253)
To meet the quality control requirements for hematology, the laboratory
must comply with the applicable requirements in paragraphs 10-1 through
10-12 above and with paragraphs a through d below. All quality control activities
must be documented.
a. Cell counts performed manually using a hemocytometer must be tested in
duplicate. One control is required for each eight hours of operation.
b. For non-manual hematology testing systems, excluding coagulation, the
laboratory must include two levels of controls each eight hours of operation.
c. For all non-manual coagulation testing systems, the laboratory must include
two levels of control each eight hours of operation and each time change
in reagents occurs.
d. For manual coagulation tests:
(1) Each individual performing tests must test two levels of controls before testing patient samples and each time a change in reagents occurs.
(2) Patient and control specimens must be tested in duplicate.
10-29. Condition: Pathology (§493.1255)
The laboratory must meet the applicable quality control requirements in
paragraphs 10-1 through 10-12 above and paragraphs 10-30 through 10-32 below
for the subspecialties for which it is certified under the specialty of
pathology. All quality control activities must be documented.
10-30. Condition: Cytology (§493.1257)
To meet the quality control requirements for cytology, the laboratory must
comply with the applicable requirements in paragraphs 10-1 through 10-12
above and paragraphs a through g below. All quality control activities must
be documented.
a. The laboratory must assure that:
(1) All gynecologic smears are stained using a Papanicolaou or modified Papanicolaou staining method.
(2) Effective measures are taken to prevent cross-contamination between gynecologic and nongynecologic specimens during the staining process.
(3) Nongynecologic specimens that have a high potential for cross-contamination are stained separately from other nongynecologic specimens, and the stains are filtered or changed following staining.
(4) Diagnostic interpretations are not reported on unsatisfactory smears.
(5) All cytology slide preparations are evaluated on the premises of a laboratory certified to conduct testing in the subspecialty of cytology.
b. The laboratory is responsible for ensuring that:
(1) Each individual evaluating cytology preparations by nonautomated microscopic technique examines no more than 100 slides (one patient per slide, gynecologic or nongynecologic, or both) in a 24 hour period, irrespective of the site or laboratory. This limit represents an absolute maximum number of slides and is not to be employed as a performance target for each individual. Previously examined negative, reactive, reparative, atypical, premalignant or malignant gynecologic cases as defined in paragraph c(1) below, previously examined nongynecologic cytology preparations, and tissue pathology slides examined by a technical supervisor qualified under paragraph 11-25b or 11-25d are not included in the 100 slide limit. (For this chapter, all references to technical supervisor refer to individuals qualified under paragraph 11-25b or 11-25d).
(2) For purposes of workload calculations, each slide preparation (gynecologic and nongyneco-logic) made using automated, semi-automated, or other liquid-based slide preparatory techniques which result in cell dispersion over one-half or less of the total available slide area and which is examined by nonautomated microscopic technique counts as one-half slide.
(3) Records are maintained of the total number of slides examined by each individual during each 24 hour period, irrespective of the site or laboratory, and the number of hours each individual spends examining slides in the 24 hour period.
(a) The maximum number of 100 slides described in paragraph b(1) above is examined in no less than an 8 hour workday.
(b) For the purposes of establishing workload limits for individuals examining slides by nonautomated microscopic technique on other than an 8 hour workday basis (includes full-time employees with duties other than slide examination and part-time employees), a period of 8 hours must be used to prorate the number of slides that may be examined. Use the formula:
No. of hours examining slides x 1008
to determine maximum slide volume to be examined.
c. The individual qualified under paragraph 11-25b or 11-25d who provides
technical supervision of cytology must ensure that:
(1) All gynecologic smears interpreted to be showing reactive or reparative changes, atypical squamous or glandular cells of undetermined significance, or to be in the premalignant (dysplasia, cervical intraepithelial neoplasia or all squamous intraepithelial lesions including human papillomavirus-associated changes) or malignant category are confirmed by a technical supervisor in cytology. The report must be signed to reflect the review or, if a computer report is generated with signature, it must reflect an electronic signature authorized by the technical supervisor in cytology.
(2) All nongynecologic cytologic preparations are reviewed by the technical supervisor in cytology. The report must be signed to reflect technical supervisory review or, if a computer report is generated with signature, it must reflect an electronic signature authorized by the technical supervisor.
(3) The slide examination performance of each cytotechnologist is evaluated and documented, including performance evaluation through the re-examination of normal and negative cases and feedback on the reactive, reparative, atypical, malignant or premalignant cases as defined in paragraph c(1) above.
(4) A maximum number of slides, not to exceed the maximum workload limit described in paragraph b above is established by the technical supervisor for each individual examining slide preparations by nonautomated microscopic technique.
(a) The actual workload limit must be documented for each individual and established in accordance with the individual's capability based on the performance evaluation as described in paragraph c(3) above.
(b) Records are available to document that each individual's workload limit is reassessed at least every 6 months and adjusted when necessary.
d. The laboratory must establish and follow a program designed to detect
errors in the performance of cytologic examinations and the reporting of
results.
(1) The laboratory must establish a program that includes a review of slides from at least 10 percent of the gynecologic cases interpreted to be negative for reactive, reparative, atypical, premalignant or malignant conditions as defined in paragraph c(1) above that are examined by each individual not qualified in paragraph 11-25b or 11-25d. This review must be done by a technical supervisor in cytology, a cytology general supervisor qualified under paragraph 11-35, or a cytotechnologist qualified under paragraph 11-38 who has the experience specified in paragraph 11-35b.
(a) The review must include negative cases selected at random from the total caseload and from patients or groups of patients that are identified as having a high probability of developing cervical cancer, based on available patient information.
(b) Records of initial examinations and rescreening results must be available.
(c) The review must be completed before reporting patient results on those cases selected.
(2) The laboratory must compare clinical information, when available, with cytology reports and must compare all malignant and premalignant (as defined in paragraph c(1) above) gynecology reports with the histopathology report, if available in the laboratory (either on-site or in storage), and determine the causes of any discrepancies.
(3) For each patient with a current high grade or above intraepithelial lesion (moderate dysplasia or CIN-2 or above), the laboratory must review all normal or negative gynecologic specimens received within the previous five years, if available in the laboratory (either on-site or in storage). If significant discrepancies are found that would affect patient care, the laboratory must notify the patient's physician and issue an amended report.
(4) The laboratory must establish and document an annual statistical evaluation of the number of cytology cases examined, number of specimens processed by specimen type, volume of patient cases reported by diagnosis (including the number reported as unsatisfactory for diagnostic interpretation), number of gynecologic cases where cytology and available histology are discrepant, the number of gynecologic cases where any rescreen of a normal or negative specimen results in reclassification as malignant or premalignant, as defined in paragraph c(1) above, and the number of gynecologic cases for which histology results were unavailable to compare with malignant or premalignant cytology cases as defined in paragraph c(1) above.
(5) The laboratory must evaluate the case reviews of each individual examining slides against the laboratory's overall statistical values, document any discrepancies, including reasons for the deviation, and document corrective action, if appropriate.
e. The laboratory report must:
(1) Clearly distinguish specimens or smears, or both, that are unsatisfactory for diagnostic interpretation.
(2) Contain narrative descriptive nomenclature for all results.
f. Corrected reports issued by the laboratory must indicate the basis for
correction.
g. The laboratory must retain all slide preparations for five years from
the date of examination, or slides may be loaned to proficiency testing
programs, in lieu of maintaining them for this time period, provided the
laboratory receives written acknowledgment of the receipt of slides by the
proficiency testing program and maintains the acknowledgment to document
the loan of such slides. Documentation for slides loaned or referred for
purposes other than proficiency testing must also be maintained. All slides
must be retrievable upon request.
10-31. Condition: Histopathology (§493.1259)
To meet the quality control requirements for histopathology, a laboratory
must comply with the applicable requirements in paragraphs 10-1 through
10-12 above and paragraphs a through e below. All quality control activities
must be documented.
a. A control slide of known reactivity must be included with each slide
or group of slides for differential or special stains. Reaction(s) of the
control slide with each special stain must be documented.
b. The laboratory must retain stained slides at least 10 years from the
date of examination and retain specimen blocks at least 2 years from the
date of examination.
c. The laboratory must retain remnants of tissue specimens in a manner that
assures proper preservation of the tissue specimens until the portions submitted
for microscopic examination have been examined and a diagnosis made by an
individual qualified under paragraph 11-25b or paragraph 11-25e(1). In addition,
an individual who meets the requirements of paragraph 11-25b, paragraph
11-25e(1) or paragraph 11-25f(1), may examine and provide reports for specimens
for skin pathology; an individual meeting the requirements of paragraph
11-25b or paragraph 11-25g(1) may examine and provide reports for ophthalmic
pathology; an individual meeting the requirements of paragraph 11-25b or
paragraph 11-25h may examine and provide reports for oral pathology specimens.
d. All tissue pathology reports must be signed by an individual qualified
as specified in paragraph c above. If a computer report is generated with
an electronic signature, it must be authorized by the individual qualified
as specified in paragraph c above.
e. The laboratory must utilize acceptable terminology of a recognized system
of disease nomenclature in reporting results.
10-32. Condition: Oral pathology (§493.1261)
To meet the quality control requirements for oral pathology, the laboratory
must comply with the applicable requirements in paragraphs 10-1 through
10-12 above and paragraph 10-31 above. All quality control activities must
be documented.
10-33. Condition: Radiobioassay (§493.1263)
To meet quality control requirements for radiobioassay, the laboratory must
comply with the applicable requirements of paragraphs 10-1 through 10-12
above. All quality control activities must be documented.
10-34. Condition: Histocompatibility (§493.1265)
In addition to meeting the applicable requirements for general quality control
in paragraphs 10-1 through 10-12 above, for quality control for general
immunology in paragraph 10-22 above and for immuno-hematology in paragraph
10-36 below, the laboratory must comply with the applicable requirements
in paragraphs a through d below. All quality control activities must be
documented.
a. For renal allotransplantation, the laboratory must meet the following
requirements:
(1) The laboratory must have available and follow criteria for:
(a) Selecting appropriate patient serum samples for crossmatching.
(b) The technique used in crossmatching.
(c) Preparation of donor lymphocytes for crossmatching.
(d) Reporting crossmatch results.
(2) The laboratory must:
(a) Have available results of final crossmatches before an organ or tissue is transplanted.
(b) Make a reasonable attempt and document efforts to have available serum specimens for all potential transplant recipients at initial typing, for periodic screening, for pre-transplantation crossmatch and following sensitizing events, such as transfusion and transplant loss.
(3) The laboratory's storage and maintenance of both recipient sera and reagents must:
(a) Be at an acceptable temperature range for sera and components.
(b) Use a temperature alarm system and have an emergency plan for alternate storage.
(c) Ensure that all specimens are properly identified and easily retrievable.
(4) The laboratory's reagent typing sera inventory (applicable only to locally constructed trays) must indicate source, bleeding date and identification number, and volume remaining.
(5) The laboratory must properly label and store cells, complement, buffer, dyes, etc..
(6) The laboratory must:
(a) HLA type all potential transplant recipients.
(b) Type cells from organ donors referred to the laboratory.
(c) Have available and follow a policy that establishes when antigen redefinition and retyping are required.
(7) The laboratory must have available and follow criteria for:
(a) The preparation of lymphocytes for HLA-A, B and DR typing.
(b) Selecting typing reagents, whether locally or commercially prepared.
(c) The assignment of HLA antigens.
(d) Assuring that reagents used for typing recipients and donors are adequate to define all major and International Workshop HLA-A, B and DR specificities for which reagents are readily available.
(8) The laboratory must:
(a) Screen potential transplant recipient sera for preformed HLA-A and B antibodies with a suitable lymphocyte panel on sera collected at the time of the recipient's initial HLA typing, and thereafter, following sensitizing events and upon request.
(b) Use a suitable cell panel for screening patient sera (antibody screen), a screen that contains all the major HLA specificities and common splits. If the laboratory does not use commercial panels, it must maintain a list of individuals for fresh panel bleeding. If the laboratory uses frozen panels, it must have a suitable storage system.
(9) The laboratory must check:
(a) Each typing tray using positive control sera, negative control sera, and positive controls for specific cell types when applicable (i.e., T cells, B cells, and monocytes).
(b) Each compatibility test (i.e. mixed lymphocyte cultures, homozygous typing cells or DNA analysis) and typing for disease associated antigens using controls to monitor the test components and each phase of the test system to ensure an acceptable performance level.
(10) Compatibility testing for cellularly-defined antigens must utilize techniques such as the mixed lymphocyte culture test, homozygous typing cells or DNA analysis.
(11) If the laboratory reports the recipient's or donor's, or both, ABO blood group and D(Rho) typing, the testing must be performed in accordance with paragraph 10-36.
(12) If the laboratory utilizes immunologic reagents (such as antibodies or complement) to remove contaminating cells during the isolation of lymphocytes or lymphocyte subsets, the efficacy of the methods must be verified with appropriate quality control procedures.
(13) At least once each month, the laboratory must have each individual performing tests evaluate a previously tested specimen as an unknown to verify his or her ability to reproduce test results. Records of the results for each individual must be maintained.
(14) The laboratory must participate in at least one national or regional cell exchange program, if available, or develop an exchange system with another laboratory in order to validate interlaboratory reproducibility.
b. If the laboratory performs histocompatibility testing for:
(1) Transfusions and other non-renal transplantation, excluding bone marrow and living transplants, all the requirements specified in this chapter, as applicable, except for the performance of mixed lymphocyte cultures, must be met.
(2) Bone marrow transplantation, all the requirements specified in this chapter, including the performance of mixed lymphocyte cultures or other augmented testing to evaluate class II compatibility, must be met.
(3) Non-renal solid organ transplantation, the results of final crossmatches must be available before transplantation when the recipient has demonstrated presensitization by prior serum screening except for emergency situations. The laboratory must document the circumstances, if known, under which emergency transplants are performed, and records must reflect any information concerning the transplant provided to the laboratory by the patient's physician.
c. Laboratories performing HLA typing for disease-associated studies must
meet all the requirements specified in this chapter except for the performance
of mixed lymphocyte cultures, antibody screening and crossmatching.
d. For laboratories performing organ donor HIV testing the requirements
of paragraph 10-22 for the transfusion of blood and blood products must
be met.
10-35. Condition: Clinical cytogenetics (§493.1267)
To meet the quality control requirements for clinical cytogenetics, the
laboratory must comply with the applicable requirements of paragraphs 10-1
through 10-12 above and with paragraphs a through d below. All quality control
activities must be documented.
a. When determination of sex is performed by X and Y chromatin counts, these
counts must be based on an examination of an adequate number of cells. Confirmatory
testing such as full chromosome analysis must be performed for all atypical
results.
b. The laboratory must have records that reflect the media used and document
the reactions observed, number of cells counted, the number of cells karyotyped,
the number of chromosomes counted for each metaphase spread, and the quality
of the banding; that the resolution is sufficient to support the reported
results; and that an adequate number of karyotypes are prepared for each
patient.
c. The laboratory also must have policies and procedures for assuring an
accurate and reliable patient sample identification during the process of
accessioning, cell preparation, photographing or other image reproduction
technique, and photographic printing, and storage and reporting of results
or photographs.
d. The laboratory report must include the summary and interpretation of
the observations and number of cells counted and analyzed and the use of
appropriate nomenclature.
10-36. Condition: Immunohematology (§493.1269)
To meet the quality control requirements for immunohematology, the laboratory
must comply with the applicable requirements in paragraphs 10-1 through
10-12 above and with paragraphs a through d below. All quality control activities
must be documented.
a. The laboratory must perform ABO group and D(Rho) typing, unexpected antibody
detection, antibody identification and compatibility testing in accordance
with manufacturer's instructions, if provided, and as applicable, with 21
CFR Part 606 (with the exception of 21 CFR 606.20a, Personnel) and 21 CFR
Part 640 et seq.
b. The laboratory must perform ABO group by concurrently testing unknown
red cells with anti-A and anti-B grouping reagents. For confirmation of
ABO group, the unknown serum must be tested with known A1 and B red cells.
c. The laboratory must determine the D(Rho) type by testing unknown red
cells with anti-D (anti-Rho) blood grouping reagent.
d. If required in the manufacturer's package insert for anti-D reagents,
the laboratory must employ a control system capable of detecting false positive
D(Rho) test results.
10-37. Condition: Transfusion services and blood banking (§493.1271)
All facilities operating under the Defense Health Programs, providing services
for the transfusion of blood and blood products, must be under the adequate
control and supervision of a laboratory director meeting the qualifications
in Chapter 11. Technical supervision in immunohematology may be provided
by an individual meeting the qualifications for technical supervisor in
paragraph 11-25k. The facility must ensure that there are facilities for
procurement, safekeeping and transfusion of blood and blood products and
blood products must be available to meet the needs of the physicians responsible
for the diagnosis, management, and treatment of patients. The facility meets
this condition by complying with the standards in paragraphs 10-38 through
10-43 below. All quality control activities must be documented.
10-38. Standard: Immunohematological collection, processing, dating periods,
labeling, and distribution of blood and blood products (§493.1273)
To meet the quality control requirements for immunohematology, the laboratory
must comply with the applicable requirements in paragraphs 10-1 through
10-12 above and with paragraphs a through d below. All quality control activities
must be documented.
a. Blood and blood product collection, processing, and distribution must
comply with 21 CFR Part 640 and 21 CFR Part 606, and the testing laboratory
must meet the applicable requirements of this document.
b. Dating periods for blood and blood products must conform to 21 CFR Part
610.53.
c. Labeling of blood and blood products must conform to 21 CFR Part 606,
Subpart G.
d. Policies to ensure positive identification of a blood or blood product
recipient must be established, documented, and followed.
10-39. Standard: Blood and blood products storage facilities (§493.1275)
a. The blood and blood products must be stored under appropriate conditions,
which include an adequate temperature alarm system that is regularly inspected.
(1) An audible alarm system must monitor proper blood and blood product storage temperature over a 24 hour period.
(2) Inspections of the alarm system must be documented.
b. If blood is stored or maintained for transfusion outside of a monitored
refrigerator, the facility must ensure and document that storage conditions,
including temperature, are appropriate to prevent deterioration of the blood
or
blood product.
10-40. Standard: Arrangement for services (§493.1277)
In the case of services provided outside the blood bank, the facility must
have an agreement reviewed and approved by the director that governs the
procurement, transfer and availability of blood and blood products.
10-41. Standard: Provision of testing (§493.1281)
All facilities operating under the Defense Health Programs, must provide
prompt ABO blood group, D(Rho) type, unexpected antibody detection and compatibility
testing in accordance with paragraph 10-36 above and for laboratory investigation
of transfusion reactions, either through the facility or under arrangement
with an approved facility on a continuous basis, under the supervision of
a pathologist or other doctor of medicine or osteopathy meeting the qualifications
of paragraph 11-25b or paragraph 11-25k below.
10-42. Standard: Retention of samples of transfused blood (§493.1283)
According to the facility's established procedures, samples of each unit
of transfused blood must be retained for further testing in the event of
reactions. The facility must promptly dispose of blood not retained for
further testing that has passed its expiration date.
10-43. Standard: Investigation of transfusion reactions (§493.1285)
All facilities operating under the Defense Health Program, according to
its established procedures, must promptly investigate all transfusion reactions
occurring in all facilities for which it has investigational responsibility
and make recommendations to the medical staff regarding improvements in
transfusion procedures. The facility must document that all necessary remedial
actions are taken to prevent future recurrences of transfusion reactions
and that all policies and procedures are reviewed to assure that they are
adequate to ensure the safety of individuals being transfused within the
facility.
Return to Table of Contents